Strainsert Force Calibration Service
Strainsert’s calibration facilities enable Strainsert to certify all our
manufactured products at our manufacturing site up to a 500,000 pound capacity.
All Strainsert force measurements are traceable to the National Institute of
Standards and Technology (NIST) and are performed by a highly-trained and
experienced staff in a controlled environment. Strainsert calibration services
include:
- Factory-site Calibrations to 500,000-lbs.
(Calibration capabilities to 4,000,000-lbs. are available in partnership with
a local calibration facility)
- NIST Traceability
- Detailed Calibration Reports
- Tension and/or Compression Calibrations
- Compliance with ANSI/NCSL Z540-1 and MIL-STD-45662A Specifications (Optional)
- Complete System Calibrations Available
- Special Overload / Proof Load and Additional Testing
Standard Strainsert Calibrations:
All of Strainsert’s standard calibration procedures result in a customer
report that includes the product model and serial number, rated capacity, test
load(s) applied, output sensitivity recorded at each test load, and a statement
certifying traceability of the measurements.
Stainsert offers several different calibration procedures and corresponding
reports based on the customer’s requirements and product design criteria.
Available Standard Strainsert Calibration Procedures:
| Procedure |
Bolts |
Studs |
Single Bridge Load Cells |
Dual Bridge Load Cells |
Tension Links |
Single Axis Clevis Pins |
Dual Bridge Clevis Pins |
Clevis Pins Bi-Axial |
| PL |
|
|
|
|
|
|
|
|
| CAL2-LO |
|
|
|
|
|
|
|
|
| CAL2-LU |
|
|
|
|
|
|
|
|
| CAL5-LO |
* |
* |
|
|
|
* |
|
|
| CAL5-LU |
|
|
* |
* |
* |
|
|
|
| CALIN-LO(**) |
|
|
|
|
|
|
|
|
| CALIN-LU(**) |
|
|
|
|
|
|
|
|
| CALSC-LO(**) |
|
|
|
|
|
|
|
|
| CALSC-LU(**) |
|
|
|
|
|
|
|
|
| CALXX-LOD |
|
|
|
|
|
|
* |
|
| CALXX-LUD |
|
|
|
|
|
|
|
|
| CALXX-LOBD |
|
|
|
|
|
|
|
* |
| CALXX-LUBD |
|
|
|
|
|
|
|
|
| ANSI/NCSL Z540-1 |
|
|
|
|
|
|
|
|
| MIL-STD-45662A |
|
|
|
|
|
|
|
|
*: Indicates the standard calibration included in the purchase price.
**: System calibration is included when transducer and indicator/signal
conditioner are purchased together as a system.
XX: Specify 2, 5, IN, or SC
CAL2: Calibration with respect to a straight line with a fixed endpoint. (Please see Cal2 Trimmed drawing below)
CAL5: Calibration with repect to the best straight line through zero. (Please see Cal5 Untrimmed drawing below)
CALIN: System calibration with an indicator readout.
CALSC: System calibration with an analog signal conditioner.
Cal2 Trimmed and Cal5 Untrimmed Load Cell Calibration |
Proof Loading Calibration (PL):
A Proof Load Calibration reports the output at Full Load. Output data is reported in units of mv/V.
Loading Only Calibration (LO):
A Loading Only Calibration reports the output of the Force Sensor at each
step of a 5 step calibration run. Force Loads are applied in 5 equal steps in
ascending direction from (No-Load) to (Full-Load), then back to (No-Load) in one
step. A total of three runs are reported.
The force sensor output is reported in mv/V. Calibration Analysis is based
on percentage of Full Scale Signal, and provides:
- Non-Linearity, Compared to Best Fitting or Predetermined Straight Line Signal
- Repetition of Loading Cycles, Zero-Load, Max. Load readings.
- Deviations of End Point , or Maximum Load, reading from Straight Line Signal.
For all functions, maximum deviation figures are reported, without averaging.
Strainsert Force Sensing Clevis Pins and Bolts are provided with a one
direction (LO) calibration at no extra charge.
Loading and Unloading Calibration (LU):
A Loading and Unloading Calibration reports the output of the Force Sensor at
each step of a 10 point calibration run. Force Loads are applied from No-Load
to Full-Load in five equal steps, then back to No-Load in five equal steps. A
total of three runs are reported.
The force sensor output is reported in mv/V. Calibration Analysis is based
on percentage of Full Scale Signal, and provides:
- Non-Linearity, Compared to Best Fitting or Predetermined Straight Line Signal
- Repetition of Loading Cycles, Zero-Load, Max. Load readings.
- Deviations of End Point , or Maximum Load, reading from Straight Line Signal.
- Hysteresis
For all functions, maximum deviation figures are reported, without averaging.
Strainsert Flat Load Cells, Fatigue Rated Flat Load Cells, and Tension Links
are provided with a one direction (LU) calibration at no extra charge.
Loading Only Calibration Dual Bridge (LOD):
Identical to the (LO) calibration for each indepedent bridge of a dual bridge
transducer.
Strainsert Dual-Bridge Load Sensing Clevis Pins and Bolts are provided with one
direction (LOD) calibration at no additional cost.
Loading and Unloading Calibration Dual Bridge (LUD):
Identical to the (LU) calibration for each independent bridge of a
dual-bridge transducer.
Strainsert Dual-Bridge Flat Load Cells, Dual Bridge Fatigue Rated Flat Load
Cells, Dual Bridge Tension Links are provided with one direction (LUD)
calibrations at no additional cost.
Loading Only Calibration Bi-Directional (LOBD):
(LOBD) calibration loads are applied from (No-Load) to (Full-Load) in five
equal steps, then back to (No-Load) in one step.
Two runs are recorded for each of the loading directions.
The transducer signal is reported in mV/V.
Calibration Analysis is based on %FS Signal, and provides:
- Non-Linearity, Compared to Best Fitting or Predetermined Straight Line Signal
- Repetition of Loading Cycles, Zero-Load, Max. Load readings.
- Deviations of End Point , or Maximum Load, reading from Straight Line Signal.
- Detailed Cross-Talk readings induced in the bridge circuit of the loaded direction.
For all functions the maximum deviation figures are reported without
averaging.
Strainsert Bi-Directional Load Sensing Clevis Pins and Bolts are provided with one (LOBD)
calibration for each primary loading direction at no additional cost. Cross-talk correction
algorithms are also provided as required.
Loading and Unloading Calibration Bi-Directional (LUBD):
(LUBD) calibration loads are applied from (No-Load) to (Full-Load) in five
equal steps, then back to (No-Load) in five equal steps. Two complete runs
are made for each of the two loading directions.
The transducer signal is reported in mV/V.
Calibration Analysis is based on %FS Signal, and provides:
- Non-Linearity, Compared to Best Fitting or Predetermined Straight Line Signal
- Repetition of Loading Cycles, Zero-Load, Max. Load readings.
- Deviations of End Point , or Maximum Load, reading from Straight Line Signal.
- Hysteresis
- Detailed Cross-Talk readings induced in the bridge circuit of the unloaded direction.
For all functions the maximum deviation figures are reported without
averaging.
Cross-talk correction algorithms are also provided as required.
|
Strainsert - for superior internally gaged force transducers.
Strainsert stands for:
- Product Quality
- Knowledgeable Technical Staff
- Standard and Custom Designs
- Customer Service
- Comprehensive Testing
Contact Strainsert
For information on load pins, force sensing bolts, load cells, tension links, force measurement systems or our high quality custom products, Contact Us for further assistance.
|
|